Trumouse Lite: Installation
Site Requirements
Section titled “Site Requirements”- Footprint: approximately 660 mm × 660 mm; height approximately 935 mm. Stationary, floor-standing installation; not for use while moving.
- Operating temperature: 20 °C to 26 °C; around 23 °C is recommended, as for mouse housing.
- Relative humidity: 30 ~ 70 %RH; Non-condensing.
- Indoors, away from light.
- Quiet, with no noticeable or sudden noise and no people walking around.
Assembly and Wiring
Section titled “Assembly and Wiring”See Figure “System topology” for the system connections.
All devices are connected to the switch by network cable (see Network Connections). Use shielded network cable; Category 5e or higher is recommended. Each Reader Array block has 2 network ports. Keep all of them connected by network cable (2 cables for the Trumouse Lite). Each block also has 2 DC power inputs (2 for the Trumouse Lite). Use the supplied 1-to-2 DC splitter cable to connect all inputs of a block to one 5 V DC supply. The network configuration is set up by Murinova before delivery; do not change it.
The DC power connectors of the components look identical, so there is a risk of connecting the wrong supply (see Electrical Safety). When wiring, connect each component only to a supply of its rated voltage, as listed below; do not swap them:
Table: Trumouse Lite rated supply of each component
| Component | Rated supply |
|---|---|
| Reader Array floor platform | 5 V DC |
| Top Panel (White Light, Infrared Light, environmental sensor, Light Control Module and Exhaust Fan) | 24 V DC |
| Industrial Camera | External 12 V / 24 V DC supply, or powered together with the Top Panel lighting |
| Compute Unit | Use the dedicated power adapter supplied with it |